XZB-W No partial
discharge Series Resonant Test System is composed of no partial discharge
variable-frequency power source, no partial discharge Excitation transformer,
no partial discharge electric reactor and no partial discharge Coupling
capacitor voltage divider. The capacitor and electric reactor of test object
constitute the connection of series resonance, voltage divider is connected
with tested object in parallel, as the series resonance voltage on test object
and Overvoltage protection signal. The frequency modulated power output is
coupled to the series resonant circuit by the excitation transformer, providing
series resonant excitation power. Frequency conversion series resonance test
device adopts series resonance theory, using excitation transformer to excites
series resonance circuit, adjust output frequency of frequency conversion
controller, and Make the circuit inductance L and sample C series resonance,
the resonant voltage is the voltage added to tested equipment.
1. HUAYI No partial discharge Series Resonant
Test System adopts power amplifiers imported from Japan, adopts ABB control
technology, can monitor the power factor, quality factor and test data
such as current, voltage and frequency in real time, has perfect protection
function and stable performance such as over voltage, over zero voltage,
under voltage, flash fall, etc.
2. No partial discharge reactor coil adopts single or multi strands winding enameled wires with withstand high temperature (200℃), with hollow core cake structure, and equipped with multi oil channels, the end of each coil is adopted pressure equalizing method to make sure no partial discharge and lower temperature. Special air bag is installed internally to ensure the space for thermal expansion and cold shrinkage of oil, meet the rated working requirements.
3. No partial discharge exciting transformer adopts winding cooper conductor and oil immersed cooling iron structure, electrostatic shielding layer is set between high and low voltage windings and iron core, which is excitation change and isolation change, so that it has enough electrical and mechanical strength to ensure no partial discharge and corona discharge during rated operation.
1. Series resonance variable frequency power technology parameters
Variable frequency power
Quantity 1
Project
Parameters
MAX power output
300kW(active power)
Input power supply
3 phase AC380V+15% 50Hz
Rated output voltage
Single phase 0~350V adjustable
Output voltage waveform
Sine wave, total harmonic THD% less than 1%
Output voltage instability
<1%( input voltage instability within 5 min less than 2%)
Rated output current
857A
Output frequency range
20~300Hz
Noise level
≤85dB
Type of cooling
Forced air cooling
Weight / dimensions
1500kg/1800×800×1800mm(L×W×H)
2. Series harmonic excitation transformer technology parameters
Excitation transformer
Quantity 1
Project
Parameters
Rated capacity
300kVA
Rated voltage of low voltage winding
2×300、350、400V
Maximum input current at low voltage side
857A
Rated voltage of high voltage winding
2×10kV ;2×35kV
Insulation level
Low voltage winding to ground:5kV/1min
High voltage winding to ground
1.2U0/1min
Rated frequency
45~300Hz
Magnitude of partial discharge
≤10pC
Impedance voltage
≤5%
Noise level
≤65dB
Type of cooling
ONAN
3. Series resonant compensation reactor technology parameters
4. Series resonant capacitor voltage divider technology
parameters
Compensation reactor
Quantity 4, parameters for single as below
Project
Parameters
Rated voltage
20kV
Inductance value
6.0 H(±10%)
Rated current
5A
Rated power
100H
Local discharge at rated voltage
≤10pC
Continuous working time
Continuous working 90min at rated current
Insulation level
1.2Un/1min
Temperature rise
≤ 65K
Weight
0.5T
Dimensions
600×1600×600mm
Capacitor voltage divider
Quantity 1(for partial discharge test)
Project
Parameters
Rated voltage
100kV
Capacity
C1=300pF
Working frequency
30~300Hz
Measurement error
≤1.5%
Insulation level
110kV/1min
Low voltage arm ratio
1000:1
Overall partial discharge of voltage divider
≤10pC
Rated voltage
100kV
Error
≤1.5%
Magnitude of partial discharge
≤10pC
Serial Number |
Name | Amount |
1 | Mainframe | one |
2 |
Specification |
one |
3 |
Inspection Report |
one |
4 |
Certification |
one |
Your Name: | |
E-mail: | |
Phone: | |
Title: | |